ATTENTION –
High and low temperature alarms set
up in the cabinets’ controller (including
EAL alarms) must be accompanied by
additional redundant independent
external alarms to ensure maximum item
safety.
ELd – Setting the delay of the external lower alarm limit [min.]
Press and hold + for more than 3 seconds
Press to proceed to “EAL”
Press to select “EAL”. “EHL” is now shown in the display
Press several times until “ELd” is shown in the display
Press to select “ELd”. The delay of the external lower alarm limit is now shown in the display
Press or to set the desired value for the delay of the lower alarm limit
Press to conrm the set value
– The delay of the external lower alarm limit is now set, proceed to other parameters by pressing
, and then navigate
by using or
.
Leave the user menu by pressing several times until the cabinet temperature is shown in the display
EHd – Setting the delay of the external upper alarm limit [min.]
Press and hold + for more than 3 seconds
Press to proceed to “EAL”
Press to select “EAL”. “EHL” is now shown in the display
Press several times until “EHd” is shown in the display
Press to select “EHd”. The external delay of the upper alarm limit is now shown in the display
Press or to set the desired value for the external delay of the upper alarm limit
Press to conrm the set value
– The delay of the external upper alarm limit is now set, proceed to other parameters by pressing
, and then
navigate by using or
Leave the user menu by pressing several times until the cabinet temperature is shown in the display
External high alarm delay
External low alarm delay