18
When setting up in an ordinary scenario that is not subject to
regulations for EN 60079-15 zone 2
The appliance may be connected in accordance with applicable
local heavy current regulations.
Note that there are special regulations for products that
are in accordance with EN 60079-15 zone 2 and EN 60079-
14: Explosive atmospheres – Electrical installations design,
selection and erection.
The appliance has been manufactured in accordance with EN
60079-15: Electrical apparatus for explosive gas atmospheres –
Part 15: Type of protection II 3G Ex ec nC ic IIB Tx Gc. Zone 2 is the
applicable zone.
If the appliance is to be installed in a zone 2 environment,
specialist personnel should perform the installation, or be
consulted beforehand, in order to ensure that the appliance is
installed in compliance with the guidelines currently contained in
the standard.
The cabinet is intended for connection to alternating current. The
connection values for voltage (V) and frequency (Hz) are given on
the type/number plate.
BioCompact II 210, 310, 210/210, 310/210, 410
The mains terminal is accessed via the preload cover on the back
of the cabinet. Unscrew the preload cover to access the mains
supply terminal. Ensure that the preload cover is reinstalled after
plugging in the power cord. The leaf spring in the preload cover
must engage and preload the plug of the cord as shown in the
illustrations below.
BioCompact II 610
The power cord is plugged in the terminal box on the back of the
cabinet. The plug is then xated in place by the hanger that is built
into the terminal box.
Please note that the hanger should be tted tightly around the
plug, as shown.
In all cases, ensure that the mains plug is seated completely in the
terminal on the cabinet.
The appliance must be connected to the external power supply
using a suitable device which mechanically prevents the plug and
socket from being separated unintentionally.
The connection must be labelled:
”DO NOT SEPARATE WHEN ENERGIZED”
Read the following part thoroughly before connecting the cabinet. Contact a qualied electrician if in doubt.
Connection to power
Fuses and similar must never be removed or replaced while
the appliance is connected to a power source. The electrical
terminal box must never be opened while the appliance is
connected to a power source.
The compressor starting equipment must never be
dismantled while the appliance is connected to a power
source.
Whenever electrical components are dismantled or
replaced, the appliance must be moved to an area in
which there is no risk of ignition caused by the electrical
components or gases contained in the appliance.
Never use the cabinet if the plug is damaged. The cabinet
should be examined by a Gram Commercial service
technician in such cases.
When setting up in an ordinary scenario that is not
subject to regulations for Zone 2: The appliance may be
connected in accordance with applicable local heavy current
regulations.
In both cases
Use a three-wire plug, if the power outlet is intended for a
three-wire plug, a the lead in green/yellow insulation should
be connected to the ground terminal.
Power must be connected via a wall socket. The wall socket
should be easily accessible.
All earthing requirements stipulated by the local electricity
authorities must be observed. The cabinet plug and wall
socket should then give correct earthing. If in doubt, contact
your local supplier or authorized electrician.
ATTENTION
For Ex environments –
Special conditions for safe use may apply to
this product when installing in an EN 60079-
14 environment. Please see corresponding Ex
certicate for specications.
Technical Support –
In the event of technical diculties always contact
Gram BioLine technical support or a Gram BioLine
authorized service partner. Never dismantle the
terminal box or any other elctrical component.