EasyManuals Logo

Gram BioCompact II User Manual

Gram BioCompact II
88 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #54 background imageLoading...
Page #54 background image
54
Technical specications Data
Temperature range -25/-5°C
Ambient temperature range +10/+35°C
Software variant F52
K-Value 0.35 W/(m2*K)
ATEX marking II 3G Ex ec nC ic llB T3 Gc
ATEX certicate DTI 21ATEX0160X
Gross volume 125 litres
Net volume 104 litres
Dimensions – H x B x D 595 x 640 x 801/1001 mm
Refrigerant R134a
Refrigerant charge 70 g
Refrigeration capacity at -25°C 156 Watt
GWP – CO2e 100.1
Energy consumption –
default setpoint
1.53 kWh/24h
Heat emission 100% 159 Watt
Heat emission default setpoint 81 Watt
Nominal consumption 158 Watt
Start current 8.6 A
Sound level 39.7 dB(A)
Technical specications Data
Temperature range -25/-5°C
Ambient temperature range +10/+35°C
Software variant F52
K-Value 0.35 W/(m2*K)
ATEX marking II 3G Ex ec nC ic llB T3 Gc
ATEX certicate DTI 21ATEX0160X
Gross volume 218 litres
Net volume 189 litres
Dimensions – H x B x D 595 x 640 x 1190/1390 mm
Refrigerant R134a
Refrigerant charge 80 g
Refrigeration capacity at -25°C 216 Watt
GWP – CO2e 114.4
Energy consumption –
default setpoint
1.99 kWh/24h)
Heat emission 100% 185 Watt
Heat emission default setpoint 83 Watt
Nominal consumption 187 Watt
Start current N/A
Sound level 39.9 dB(A)
BioCompact II RF210H BioCompact II RF310H
BioCompact II RF210H
Solid door
BioCompact II RF310H
Solid door

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Gram BioCompact II and is the answer not in the manual?

Gram BioCompact II Specifications

General IconGeneral
BrandGram
ModelBioCompact II
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals