2828
b i o l ine
On/O local door alarm
Delay for local door alarm
dAd – Setting the delay of the local door alarm [min.]
Press and hold + for more than 3 seconds
Press to select “LAL”. “LHL” is now shown in the display
Press several times until “dAd” is shown in the display
Press to select “dAd”. The delay of the local door alarm is now shown in the display
Press or to set the desired value for the delay of the local door alarm
Press to conrm the set value
– The delay of the local door alarm is now congured, proceed to other parameters by pressing
, and then navigate
by using or
.
Leave the user menu by pressing several times until the cabinet temperature is shown in the display
dA – Activate/deactivate of local door alarm
Press and hold + for more than 3 seconds
Press to select “LAL”. “LHL” is now shown in the display
Press several times until “dA” is shown in the display
Press to select “dA”.
Press or to activate/deactivate the local door alarm [1 = activated/0 = deactivated]
Press to conrm the set value
– The local door alarm is now congured, proceed to other parameters by pressing , and then navigate by using or
.
Leave the user menu by pressing several times until the cabinet temperature is shown in the display
ATTENTION –
High and low temperature alarms set
up in the cabinets’ controller (including
EAL alarms) must be accompanied by
additional redundant independent
external alarms to ensure maximum item
safety.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Gram BioCompact II and is the answer not in the manual?

Gram BioCompact II Specifications

General IconGeneral
BrandGram
ModelBioCompact II
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals