Escorted/set alarm limits
ALL – Setting of escorted/set alarm limits
Press and hold +
for more than 3 seconds
Press
several times until “ALL” is shown in the display
Press to select “ALL”
Press or to select set or escorted alarm limits
Press to conrm the set value
Leave the user menu by pressing several times until
the cabinet temperature is shown in the display
“Set alarm” is xed limits working independently from the setpoint.The temperature alarm limits will remain the selected values
regardless of the setpoint being altered.
“Escorted alarm” is xed limits locked to the setpoint. The temperature alarm limits will change according to the altered
setpoint.
ATTENTION –
High and low temperature alarms set
up in the cabinets’ controller (including
EAL alarms) must be accompanied by
additional redundant independent
external alarms to ensure maximum item
safety.