Oset of the A-sensor
Press and hold + for more than 3 seconds
Press several times until “cAL” is shown in the display
Press to select “cAL”. “cA” is shown in the display
Press to select “cA”
Press or to oset the A-sensor
Press to conrm the set value
– The A-sensor is now oset, proceed to other parameters by pressing
, and then navigate by using or
Leave the user menu by pressing several times until the cabinet temperature is shown in the display
Oset of the E-sensor
Press and hold + for more than 3 seconds
Press several times until “cAL” is shown in the display
Press to select “cAL”. “cA” is shown in the display
Press until “cE” is shown in the display
Press to select “cE”
Press or to oset the E-sensor
Press to conrm the set value
– The E-sensor is now oset proceed to other parameters by pressing
, and then navigate by using or
Leave the user menu by pressing several times until the cabinet temperature is shown in the display
Oset of the F-sensor
Press and hold + for more than 3 seconds
Press several times until “cAL” is shown in the display
Press to select “cAL”. “cA” is shown in the display
Press until “cF” is shown in the display
Press to select “cF”
Press or to oset the F-sensor
Press to conrm the set value
– The F-sensor is now oset proceed to other parameters by pressing
, and then navigate by using or
Leave the user menu by pressing several times until the cabinet temperature is shown in the display