b i o l ine
Local high alarm
Local low alarm
Local alarm settings
LLL – Setting the lower alarm limit [° C]
Press and hold + for more than 3 seconds
Press to select “LAL”. “LHL” is now shown in the display
Press to proceed to “LLL”
Press to select “LLL”. The lower alarm limit is now shown in the display
Press or to set the desired value for the lower alarm limit
Press to conrm the set value
– The lower alarm limit is now set, proceed to other parameters by pressing
, and then navigate by using or .
Leave the user menu by pressing several times until the cabinet temperature is shown in the display
LHL – Setting the upper alarm limit [° C]
Press and hold + for more than 3 seconds
Press to select “LAL”. “LHL” is now shown in the display
Press to select “LHL”. The upper alarm limit is now shown in the display
Press or to set the desired value for the upper alarm limit
Press to conrm the set value
– The upper alarm limit is now set, proceed to other parameters by pressing
, and then navigate by using or
.
Leave the user menu by pressing several times until the cabinet temperature is shown in the display
ATTENTION –
High and low temperature alarms set
up in the cabinets’ controller (including
EAL alarms) must be accompanied by
additional redundant independent
external alarms to ensure maximum item
safety.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Gram BioCompact II and is the answer not in the manual?

Gram BioCompact II Specifications

General IconGeneral
BrandGram
ModelBioCompact II
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals