66
b i o l ine
BioCompact II RR210/RF210G
Technical specications Data
Temperature range +2/+20°C and -25/-5°C
Ambient temperature range +10/+35°C
Software variant K82/F52
K-Value 0.35 W/(m2*K)
ATEX marking II 3G Ex nA nC nL llB T3 Gc
ATEX certicate DTI 21ATEX0160X
Gross volume 2 x 125 litres
Net volume 2 x 104 litres
Dimensions – H x B x D 595 x 640 x 1602/1802 mm
Refrigerant R600a
Refrigerant charge 63 g
Refrigeration capacity at -10°C 154 Watt
Refrigeration capacity at -25°C 156 Watt
GWP – CO2e N/A
Energy consumption –
default setpoint
2.14 kWh/24h
Heat emission 100% 243 Watt
Heat emission default setpoint 90 Watt
Nominal consumption 239 Watt
Start current 16.2 A
Sound level 41.4 dB(A)
Technical specications Data
Temperature range +2/+20°C and -25/-5°C
Ambient temperature range +10/+32°C
Software variant K92/F52
K-Value
Solid door 0.35 W/(m2*K)
Glass door 0.49 W/(m2*K)
ATEX marking II 3G Ex nA nC nL llB T3 Gc
ATEX certicate DTI 21ATEX0160X
Gross volume 2 x 125 litres
Net volume 2 x 104 litres
Dimensions – H x B x D 595 x 640 x 1602/1802 mm
Refrigerant R600a
Refrigerant charge 63 g
Refrigeration capacity at -10°C 154 Watt
Refrigeration capacity at -25°C 156 Watt
GWP – CO2e N/A
Energy consumption –
default setpoint
2.14 kWh/24h
Heat emission 100% 243 Watt
Heat emission default setpoint 105 Watt
Nominal consumption 244 Watt
Start current 16.2 A
Sound level 41.4 dB(A)
BioCompact II RR210/RF210G
Solid door
BioCompact II RR210/RF210G
Glass door (RR)