EasyManuals Logo

Gram BioCompact II User Manual

Gram BioCompact II
88 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #73 background imageLoading...
Page #73 background image
73
b i o l ine
Technical specications Data
Temperature range +2/+20°C
Ambient temperature range +10/+38°C
Software variant K6+
K-Value 0.46 W/(m2*K)
ATEX marking II 3G Ex nA nC nL llB T6 Gc
ATEX certicate
Gross volume 583 litres
Net volume 419 litres
Dimensions – H x B x D 695 x 875 x 1874/2075 mm
Refrigerant R600a
Refrigerant charge 102 g
Refrigeration capacity at -10°C 354 Watt
GWP – CO2e N/A
Energy consumption –
default setpoint
1.9 kWh/24h
Heat emission 100% 240 Watt
Heat emission default setpoint 79 Watt
Nominal consumption 315 Watt
Start current 8 A
Sound level 44.2 dB(A)
BioCompact II RR610G
Technical specications Data
Temperature range +2/+20°C
Ambient temperature range +10/+43°C
Software variant K2+
K-Value 0.31 W/(m2*K)
ATEX marking II 3G Ex nA nC nL llB T6 Gc
ATEX certicate
Gross volume 583 litres
Net volume 419 litres
Dimensions – H x B x D 695 x 875 x 1874/2075 mm
Refrigerant R600a
Refrigerant charge 102 g
Refrigeration capacity at -10°C 354 Watt
GWP – CO2e N/A
Energy consumption –
default setpoint
1.15 kWh/24h
Heat emission 100% 205 Watt
Heat emission default setpoint 48 Watt
Nominal consumption 291 Watt
Start current 9.3 A
Sound level 44.2 dB(A)
BioCompact II RR610G
Solid door
BioCompact II RR610G
Glass door

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Gram BioCompact II and is the answer not in the manual?

Gram BioCompact II Specifications

General IconGeneral
BrandGram
ModelBioCompact II
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals