Technical specications Data
Temperature range -25/-5°C
Ambient temperature range +10/+43°C
Software variant F51
K-Value 0.31 W/(m2*K)
ATEX marking II 3G Ex nA nC nL llB T2 Gc
ATEX certicate
Gross volume 583 litres
Net volume 419 litres
Dimensions – H x B x D 695 x 875 x 1874/2075 mm
Refrigerant R404a
Refrigerant charge 220 g
Refrigeration capacity at -25°C 616 Watt
GWP – CO2e 862.85
Energy consumption –
default setpoint
4.2 kWh/24h
Heat emission 100% 583 Watt
Heat emission default setpoint 175 Watt
Nominal consumption 613 Watt
Start current 19.6 A
Sound level 48.6 dB(A)
BioCompact II RF610H
Solid door