51
b i o l ine
51
BioCompact II 210, 310, 410, 210/210, 310/210
BioCompact II RR210H
Technical specications Data
Temperature range +2/+20°C
Ambient temperature range +10/+35°C
Software variant K82
K-Value 0.35 W/(m2*K)
ATEX marking II 3G Ex ec nC ic llB T5 Gc
ATEX certicate DTI 21ATEX0160X
Gross volume 125 litres
Net volume 104 litres
Dimensions – H x B x D 595 x 640 x 801/1001mm
Refrigerant R134a
Refrigerant charge 80 g
Refrigeration capacity at -10°C 150 Watt
GWP – CO2e 114.4
Energy consumption –
default setpoint
0.68 kWh/24h
Heat emission 100% 111 Watt
Heat emission default setpoint 28 Watt
Nominal consumption 110 Watt
Start current 6.9 A
Sound level 36.5 dB(A)
BioCompact II RR210H
Solid door
Technical specications Data
Temperature range +2/+20°C
Ambient temperature range +10/+32°C
Software variant K82
K-Value 0.49 W/(m2*K)
ATEX marking II 3G Ex ec nC ic llB T5 Gc
ATEX certicate DTI 21ATEX0160X
Gross volume 125 litres
Net volume 104 litres
Dimensions – H x B x D 595 x 640 x 801/1001 mm
Refrigerant R134a
Refrigerant charge 80 g
Refrigeration capacity at -10°C 150 Watt
GWP – CO2e 114.4
Energy consumption –
default setpoint
1.22 kWh/24h
Heat emission 100% 190 Watt
Heat emission default setpoint 51 Watt
Nominal consumption 200 Watt
Start current 9.7 A
Sound level 36.5 dB(A)
BioCompact II RR210H
Glass door

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Gram BioCompact II and is the answer not in the manual?

Gram BioCompact II Specifications

General IconGeneral
BrandGram
ModelBioCompact II
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals