b i o l ine
External high alarm
External low alarm
External alarm settings
ELL – Setting the external lower alarm limit [° C]
Press and hold + for more than 3 seconds
Press several times until “EAL” is shown in the display
Press to select “EAL”. “EHL” is now shown in the display
Press to proceed to “ELL”
Press to select “ELL”. The external lower alarm limit is now shown in the display
Press or to set the desired value for the external lower alarm limit
Press to conrm the set value
– The external lower alarm limit is now set, proceed to other parameters by pressing
, and then navigate by using or
.
Leave the user menu by pressing several times until the cabinet temperature is shown in the display
EHL – Setting the external upper alarm limit [° C]
Press and hold + for more than 3 seconds
Press several times until “EAL” is shown in the display
Press to select “EAL”. “EHL” is now shown in the display
Press to select “EHL”. The external upper alarm limit is now shown in the display
Press or to set the desired value for the external upper alarm limit
Press to conrm the set value
– The external upper alarm limit is now set, proceed to other parameters by pressing
, and then navigate by using or
.
Leave the user menu by pressing several times until the cabinet temperature is shown in the display
ATTENTION –
High and low temperature alarms set
up in the cabinets’ controller (including
EAL alarms) must be accompanied by
additional redundant independent
external alarms to ensure maximum item
safety.