EasyManuals Logo

Gram BioCompact II User Manual

Gram BioCompact II
88 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #24 background imageLoading...
Page #24 background image
24
Menu Access +
Local alarm settings LAL
LHL [° C] Upper alarm limit. Code for activated alarm [A2]
LLL [° C] Lower alarm limit. Code for activated alarm [A3]
LHd [Min.] Delay of upper alarm limit
LLd [Min.] Delay of lower alarm limit
dA On/o Door alarm. Code for activated alarm [ A1]. [1=on/0=o]
dAd [Min.] Delay of door alarm
bU On/o Acoustic signal for alarm codes [ A1], [ A2] and [ A3]. [1=on/0=o]
External alarm settings EAL
EHL [° C] Upper alarm limit. Code for activated alarm [A4]
ELL [° C] Lower alarm limit. Code for activated alarm [A5]
EHd [Min.] Delay of upper alarm limit
ELd [Min.] Delay of lower alarm limit
dA On/o Door alarm. Code for activated alarm [ A1]. [1=on/0=o]
dAd [Min.] Delay of external door alarm
bU On/o Acoustic signal for external alarm codes [ A1], [ A4], [ A5]. [1=on/0=o]
Oset of sensors CAL
CA [° K] Oset setting of A-sensor. Reference sensor for the refrigeration system
CE [° K] Oset setting of E-sensor. Reference sensor for the display and alarms
CF [° K]
Oset setting of F-sensor. Reference sensor for the low temperature
protection
Electric low temperature
protection
FP
ACt On/o Activation/deactivation of low-temperature protection
tES On Test of low-temperature protection
SEt [° C] Setting of the cut-o temperature for the low-temperature protection
PrE […] Read out of the real-time temperature of the F-sensor
ALL Activation of escorted alarm limits. [FAS]=locked limits/[ESC]=follows setpoint
dEF Number of defrosts per 24 hours (4 is factory setting)
dPS Reference sensor for the display (A, E or F)
Buttons Duration Function
+
> 3 seconds Start or stop a defrost
+
> 6 seconds Activating/deactivating the keypad lock
- Shows temperature setpoint value
- Shows the highest registered temperature spike (since the last reset of alarm history)
-
Shows the lowest registered temperature spike
(since the last reset of alarm history)
+
> 3 seconds Reset alarm history
+
+
> 6 seconds Restore to factory settings
+
> 3 seconds Access to user menu and alarm settings
Walkthrough of menu
Lighting inside the cabinet – Only applicable to BioCompact ll cabinets with glass doors
The settings for the lighting inside the cabinet can be changed by pressing and holding for 3 seconds.
There are two settings: 1) Light turns on when the door is open (o when closed)
2) Light is always on

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Gram BioCompact II and is the answer not in the manual?

Gram BioCompact II Specifications

General IconGeneral
BrandGram
ModelBioCompact II
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals