BioCompact II RF410H
Technical specications Data
Temperature range -25/-5°C
Ambient temperature range +10/+35°C
Software variant F52
K-Value 0.35 W/(m2*K)
ATEX marking II 3G Ex ec nC ic llB T3 Gc
ATEX certicate DTI 21ATEX0160X
Gross volume 346 litres
Net volume 312 litres
Dimensions – H x B x D 595 x 640 x 1776/1976* mm
Refrigerant R134a
Refrigerant charge 95 g
Refrigeration capacity at -25°C 216 Watt
GWP – CO2e 135.85
Energy consumption –
default setpoint
2.66 kWh/24h
Heat emission 100% 190 Watt
Heat emission default setpoint 111 Watt
Nominal consumption 191 Watt
Start current 11.3 A
Sound level 38.6 dB(A)