32
On/O external door
External
door alarm delay alarm
dAd – Setting the delay of the external door alarm [min.]
Press and hold + for more than 3 seconds
Press to proceed to “EAL”
Press to select “EAL”. “EHL” is now shown in the display
Press several times until “dAd” is shown in the display
Press to select “dAd”. The delay of the external door alarm is now shown in the display
Press or to set the desired value for the delay of the external door alarm
Press to conrm the set value
– The delay of the external door alarm is now congured, proceed to other parameters by pressing
, and then
navigate by using
or
.
Leave the user menu by pressing several times until the cabinet temperature is shown in the display
dA – Activation/deactivation of external door alarm
Press and hold + for more than 3 seconds
Press to proceed to “EAL”
Press to select “EAL”. “EHL” is now shown in the display
Press several times until “dA” is shown in the display
Press to select “dA”
Press or to activate/deactivate the external door alarm [1 = activated/0 = deactivated]
Press to conrm the set value
– The external door alarm is now congured, proceed to other parameters by pressing
, and then
navigate by using or
Leave the user menu by pressing several times until the cabinet temperature is shown in the display
ATTENTION –
High and low temperature alarms set
up in the cabinets’ controller (including
EAL alarms) must be accompanied by
additional redundant independent
external alarms to ensure maximum item
safety.