53
b i o l ine
Technical specications Data
Temperature range +2/+20°C
Ambient temperature range +10/+35°C
Software variant K82
K-Value 0.35 W/(m2*K)
ATEX marking II 3G Ex ec nC ic llB T4 Gc
ATEX certicate DTI 21ATEX0160X
Gross volume 346 litres
Net volume 312 litres
Dimensions – H x B x D 595 x 640 x 1776/1976 mm
Refrigerant R134a
Refrigerant charge 120 g
Refrigeration capacity at -10°C 173 Watt
GWP – CO2e 171.6
Energy consumption –
default setpoint
0.96 kWh/24 h
Heat emission 100% 140 Watt
Heat emission default setpoint 40 Watt
Nominal consumption 140 Watt
Start current 8.6A
Sound level 35.1 dB(A)
BioCompact II RR410H
Solid door
BioCompact II RR410H
Technical specications Data
Temperature range +2/+20°C
Ambient temperature range +10/+32°C
Software variant K92
K-Value 0.51 W/(m2*K)
ATEX marking II 3G Ex ec nC ic llB T4 Gc
ATEX certicate DTI 21ATEX0160X
Gross volume 346 litres
Net volume 312 litres
Dimensions – H x B x D 595 x 640 x 1776/1976 mm
Refrigerant R134a
Refrigerant charge 120 g
Refrigeration capacity at -10°C 173 Watt
GWP – CO2e 171.6
Energy consumption –
default setpoint
1.58 kWh/24h
Heat emission 100% 199 Watt
Heat emission default setpoint 66 Watt
Nominal consumption 197 Watt
Start current 5.9A
Sound level 35.1 dB(A)
BioCompact II RR410H
Glass door