ATTENTION –
High and low temperature alarms set
up in the cabinets’ controller (including
EAL alarms) must be accompanied by
additional redundant independent
external alarms to ensure maximum item
safety.
Local acoustic settings
bU – Activation/deactivation of the acoustic local alarms
Press and hold + for more than 3 seconds
Press to select “LAL”. “LHL” is now shown in the display
Press several times until “bU” is shown in the display
Press to select “bU”.
Press or to activate/deactivate the local acoustic alarms [1 = activated/0 = deactivated]
Press to conrm the set value
– The local acoustic alarms is congured, proceed to other parameters by pressing
, and then navigate by using
or
.
Leave the user menu by pressing several times until the cabinet temperature is shown in the display