b i o l ine
Quick guide – BioCompact II
On/O
Press the button to turn the cabinet on. Press the button for 6 seconds to turn the cabinet o. The software version of the
cabinet will be shown when turning the cabinet on, followed by the software variant. The cabinet is ready when the temperature is
displayed. The cabinet will automatically start a defrost-cycle when turned on, and terminate it again after a system check.
Setting the temperature
Temperature adjustments are done by holding the button and pressing either or
.
Conrm the settings by letting go of the buttons.
Lighting inside the cabinet – Only applicable to BioCompact ll cabinets with glass door
The settings for the lighting inside the cabinet can be changed by pressing and holding for 3 seconds.
There are two settings: 1) Light turns on when the door is open (o when closed)
2) Light is always on
Menu Access
+
Display code and its message
Local alarm settings LAL LHL [° C] Upper temperature limit. Code for activated alarm [A2]
LLL [° C] Lower temperature limit. Code for activated alarm [A3]
LHd [min.] Delay of upper temperature limit
LLd [min.] Delay of lower temperature limit
dA On/o Door alarm. Code for activated alarm [ A1]. [1=on/0=o]
dAd [min.] Delay of door alarm
bU On/o Acoustic signal for alarm codes [ A1], [ A2] and [ A3]. [1=on/0=o]
Oset of sensors CAL CA [K] Oset setting for A-sensor. Reference sensor for the refrigeration system
CE [K] Oset setting for E-sensor. Reference sensor for the display and alarms
CF [K]
Oset setting for F-sensor. Reference sensor for the low-temperature
protection
Low-temperature protection FP ACt On/o Activation/deactivation of low-temperature protection
tES On Test of low-temperature protection
SEt [° C] Setting of the cut-o temperature for the low-temperature protection
PrE […] Read-out of the real-time temperature of the F-sensor
ALL Activation of escorted alarm limits. [FAS]=locked limits/[ESC]=follows setpoint
dEF Number of defrosts per 24 hours (4 is factory setting)
dPS Reference sensor for the display (A, E or F)
User menu and alarm settings
Temperature settings On/O
Parameter settings
Defrosting
Key pad lock is
activated
N/A
Alarms A2, A3, A4, A5

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Gram BioCompact II and is the answer not in the manual?

Gram BioCompact II Specifications

General IconGeneral
BrandGram
ModelBioCompact II
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals