Press the “Read Marker” button in the iMark software, then the iMark software read the
calibration data. And press the “Set” button in the software, then the iMark software
saved calibration data.
(リードマーカー)ボタンを押して、キャリブレーションデータを読み込みます。
NOTICE
If “Read Maker” is performed while the
calibration sheet is reflecting by the
lighting in the room, the correct
calibration may not be performed, in this
case, please shade the camera part and
then perform “Read Maker”
NOTICE
When correct calibration is performed, the value of *90 to 120 is displayed in the “New
values Delta X” and the value of *95 to 135 is displayed in the “New values Delta Y”.
(*The values are ideal values, there are varies depending on the equipment.)
If values for Delta X and Delta Y are much different than the above ideal values,
the calibration was not performed correctly. Please calibration again.