EasyManua.ls Logo

GSi TYMPSTAR 2 - Page 115

GSi TYMPSTAR 2
224 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Calibration
Step 2
Follow the instructions on the screen to
measure the ambient pressure calibra-
tion using an electronic manometer. The
manometer and LCD meter should agree
within ± 10 daPa.
Adjusting the probe Step 3
If it is necessary to adjust the probe’s
ambient pressure potentiometer:
• Loosen the cable strain relief by hand.
• Remove the two metal, and two nylon screws.
CAUTION
These screws must be reinstalled in the same positions later.
• Carefully remove the probe cover.
CAUTION
Do not disconnect the shield wire.
• Adjust the ambient potentiometer
as shown to achieve an acceptable
reading on the LCD meter.
Step 4
Proceed to the set -600/400 calibration
function or reinstall the cover and press
RETURN to return to the Main Calibra-
tion Menu.
GSI TympStar Version 1 and Version 2 Service Manual 4 - 43

Table of Contents