GDM-8245 Service Manual
60
Capacitance Calibration (Open)
GDM-8245: CL30 calibration mode
1. Enter CL30 Calibration mode.
2. Press the Auto/Man key. The secondary display will
show CL31.
3. The instrument will automatically perform an open
calibration for all capacitance ranges. After
approximately 50 seconds, the secondary display will
show CL30 again, indicating the capacitance open
calibration is complete.
Capacitance calibration (open) is complete
Resistance Calibration
±3 digits (all ranges, 19MΩ±40 digits for 20MΩ range)
Resistance Calibrator: 400Ω, 4kΩ, 40kΩ, 400kΩ, 3MΩ,
9.907MΩ, 19MΩ
GDM-8245 Range: 500Ω, 5kΩ, 50kΩ, 500kΩ, 5MΩ,
20MΩ
1. Enter CL50 Calibration mode.
2. Press Ω key to enter Resistance mode. (Range: 500 Ω)
4. Press SHIFT + REL to negate the resistance of the test
lead cables.