Calibration
65
*Press SHIFT+DCA to enter the 20A range.
The Up and Down arrow keys can be used to change the
DMM DCA range up and down.
DCA calibration is complete
ACA Calibration
GDM-8245: 500uA, 5.0mA, 50mA, 500mA, 2A, 20A
Calibrator: 200uA/70Hz, 2.0mA/70Hz, 20mA/70Hz,
200mA/70Hz, 2A/50Hz, 8A (Display 8.010A)/400Hz
1. Enter CL50 Calibration mode.
Ensure the correct current terminal is used for each
range.
2. Set the DMM to ACA mode. Press ACA. Ensure the
range is set to 500uA.
4. Check the DMM display. If the current is within
specification, go to the next calibration range.
Range: 200uA±4 digits