Calibration
65
2A
2A
2A±3
20A*
8A
8A±3
*Press SHIFT+DCA to enter the 20A range.
Note
The Up and Down arrow keys can be used to change the
DMM DCA range up and down.
DCA calibration is complete
ACA Calibration
Accepted range
± 4 digits (all ranges)
Equipment
Multimeter calibrator
Test leads
Configurations
GDM-8245: 500uA, 5.0mA, 50mA, 500mA, 2A, 20A
Calibrator: 200uA/70Hz, 2.0mA/70Hz, 20mA/70Hz,
200mA/70Hz, 2A/50Hz, 8A (Display 8.010A)/400Hz
≤2A
<20A
Verification
procedure
1. Enter CL50 Calibration mode.
Note
Ensure the correct current terminal is used for each
range.
2. Set the DMM to ACA mode. Press ACA. Ensure the
range is set to 500uA.
3. Input 200uA/70Hz.
4. Check the DMM display. If the current is within
specification, go to the next calibration range.
Range: 200uA±4 digits