Calibration
67
2. Set the DMM to mode. Press . Ensure the range
is set to 5nF.
3. Input 3.282nF capacitance.
4. Check the DMM display. If the capacitance is within
specification, go to the next calibration range.
Range: 3.282nF±3 digits
5. Press Auto/Man to re-calibrate the range. The mode
will change to CL51. Press Auto/Man again to save
when the calibration has settled. The mode will return
to CL50.
6. Repeat steps 2 to 6 for the following inputs and ranges.
The Up and Down arrow keys can be used to change the
DMM capacitance range up and down.
ACA calibration is complete
Exit Calibration Mode
1. To exit the calibration mode,
press SHIFT and the Auto/Man
key.