Calibration
67
2. Set the DMM to mode. Press . Ensure the range
is set to 5nF.
3. Input 3.282nF capacitance.
4. Check the DMM display. If the capacitance is within
specification, go to the next calibration range.
Range: 3.282nF±3 digits
5. Press Auto/Man to re-calibrate the range. The mode
will change to CL51. Press Auto/Man again to save
when the calibration has settled. The mode will return
to CL50.
6. Repeat steps 2 to 6 for the following inputs and ranges.
Range
Input
Specification
50nF
30nF
30nF±3
500nF
300nF
300nF±3
5uF
3uF
3uF±3
50uF
30uF
30uF±3
Note
The Up and Down arrow keys can be used to change the
DMM capacitance range up and down.
ACA calibration is complete
Exit Calibration Mode
Exit procedure
1. To exit the calibration mode,
press SHIFT and the Auto/Man
key.
SHIFT
AUTO/MAN
CAL