GDM-8245 Service Manual
76
1. Set the range to 5V (GDM-8245).
2. Input 4.9V/20Hz. Verify the results.
3. Repeat for the remaining ranges and frequencies.
There is no adjustment point for the 1kV range.
However, the 1kV/50Hz range can be downward
calibrated. After the downward calibration, the 1kHz
frequency can then be tested again.
ACV verification (Part 2) is complete
DCV Verification (Short)
1000V ± 7 digits, all other ranges ± 3 digits
GDM-8245 range: 500mV, 5V, 50V, 500V, 1000V
1. Set the range to 500mV (DC mV).
2. Short the terminals. Verify the results.
3. Repeat for the remaining ranges (DCV mode).
DCV Verification (short) is complete
DCV Verification (Part 1)
(Digits): +490mV±14, -490mV±14, 4.9V±14
GDM-8245: 500mV, 5V
Calibrator: -490mV, +490mV, 4.9V