Verification Log
23
VERIFICATION LOG
Print out these pages and record the results. Keep it with the instrument.
Model name
GDM-824__
Serial number
________________
Date
Year___________
Month__________
Date___________
Verified by
Name____________________________
Company/Contact_________________________
Environment
Temperature______°C
Humidity______%
Operating Voltage Verification
Item
Min limit
Result
Max limit
Pass/Fail
TP10~ GND
+14.5V
____V
15.5V
Pass
Fail
TP11~ GND
-15.5V
____V
-14.5V
Pass
Fail
TP12~ GND
+2.9V
____V
+3.3v
Pass
Fail
TP13~ GND
-3.3V
____V
-2.9V
Pass
Fail
TP14~ GND
+1.6V
____V
+2.0V
Pass
Fail
TP15~ GND
6.0V
____V
+6.4V
Pass
Fail
LED brightness Verification
Item
Pass/Fail
Primary LED
Pass
Fail
Secondary LED
Pass
Fail
Power Supply Verification
Item
Min limit
Result
Max limit
Pass/Fail
115V input
40mA
____A
50mA
Pass
Fail
230V input
17mA
____A
27mA
Pass
Fail
Resistance Verification (Short)
Item
Min limit
Result
Max limit
Pass/Fail
Short (1Ω Range)
- 2 digits
______
+ 2 digits
Pass
Fail