Verification
71
1. Set the DMM to capacitance mode. Set the range to 5nF.
Open terminals. Verify the results.
2. Repeat the procedure for all other ranges.
Capacitance verification (open) is complete
Capacitance Verification
(Digits): 1.006nF±28, 49nF±76, 490nF±71, 4.9uF±71, 49uF±71
Multimeter
calibrator/fixure
GDM-8245: 5nF, 50nF, 500nF, 5uF, 50uF
Capacitor calibrator/fixture: 1.006nF, 49nF, 490nF,
4.9uF, 49uF
1. Set the range to 5nF. Input 1.006nF. Verify the
capacitance is 1.006nF±28.
2. Repeat the procedure for the rest of the ranges.
Capacitance verification is complete
DCA Verification (Short)
GDM-8245 range: 500uA, 5.0mA, 50mA, 500mA, 2A,
20A