Verification
73
3. Repeat the procedure for the remainder of the ranges
(5.0mA~20A).
Note: Ensure the correct terminal is used for each
range.
DCA Verification is complete
ACA Verification
Accepted range
(Digits): 490uA(20Hz)<378; 490uA(2kHz)<195;
490uA(20kHz)<746
4.9mA(20Hz)<378; 4.9mA(2kHz)<195;
4.9mA(20kHz)<746
49mA(20Hz)<378; 49mA(2kHz)<195;
49mA(20kHz)<746
490mA(20Hz)<378; 490mA(400Hz)<195;
490mA(2kHz)<195
1.9A(20Hz)<153; 1.9A(400Hz)<82; 1.9A(2kHz)<82
19A*(20Hz)<153; 19A*(400Hz)<82; 19A*(2kHz)<82
Equipment
Multimeter calibrator
Test leads
Configurations
GDM-8245: 500uA, 5.0mA, 50mA, 500mA, 2A, 20A
Calibrator: 490uA/20Hz/2kHz/20kHz,
4.9mA/20Hz/2kHz/20kHz, 49mA/20Hz/2kHz/20kHz,
490mA/20Hz/400Hz/2kHz, 1.9A/20Hz/400Hz/2kHz,
19A* /20Hz/400Hz/2kHz
<2A
<20A*
*Do not input 19A for more than 1 second. Exceeding 1
second may blow the fuse.
Calibration
procedure
1. Set the range to 500uA (ACA) on the GDM-8245.
2. Input 490uA/20Hz. Verify 490uA(20Hz)<378