Calibration
61
Note
(Do not set the calibrator output to standby after REL is
pressed else the test lead resistance will not be
compensated.)
5. Input a 400Ω resistance.
6. Check the DMM display. If the resistance is within
specification, go to the next calibration range.
Range: 400Ω±3 digits
7. Press Auto/Man to re-calibrate the range. The mode
will change to CL51. Press Auto/Man again to save
when the calibration has settled. The mode will return
to CL50.
8. Repeat steps 5 to 8 for the following inputs and ranges.
Range
Input
Specification
5kΩ
4kΩ
4kΩ±3
50kΩ
40kΩ
40kΩ±3
500kΩ
400kΩ
400kΩ±3
5MΩ
3MΩ
3MΩ±3
20MΩ
9.907MΩ
9.912MΩ±3*
20MΩ
19MΩ
19MΩ±40
*For the 9.907 MΩ input, a reading of 9.912 ±3 MΩ is
required.
Note
The Up and Down arrow keys can be used to change the
DMM resistance range up and down.
Resistance calibration is completed.
DCV Calibration
Accepted range
± 3 digits (all ranges)
Equipment
Multimeter calibrator
Test leads
Configurations
GDM-8245: 500mV, 5V, 50V, 500V, 1000V
Calibrator: 400mV, 4V, 40V, 400V, 1000V