Verification
69
Resistance Verification (Short)
Accepted range
Shorting (500Ω range ±3 digits, other ranges ±2 digits)
Equipment
Test leads
Configurations
GDM-8245 Range: 500, 5k, 50k, 500k, 5M,
20M
Verification
procedure
1. Set the GDM-8245 range to 500Ω.
2. Short the test leads. Verify the resistance as 0±3 digits.
3. Repeat steps 1 and 2 for the remainder of the ranges.
Verify the resistance as 0±2 digits.
Resistance shorting verification is complete.
Resistance Verification
Accepted range
(Digits): (1Ω ±4, 100Ω±11, 490Ω±37, 4.9kΩ±36, 49kΩ±36,
490k±36, 4.9M±70, 19M±41.
Equipment
Multimeter calibrator
Test leads
Configurations
Multimeter Calibrator: 1Ω, 100, 490Ω, 4.9k, 49k,
490k, 4.9M, 19M
GDM-8245 Range: 500, 5k, 50k, 500k, 5M,
20M
Verification
procedure
1. Set the GDM-8245 range to 500Ω.
2. Input 0Ω to the DMM.