106
Parameter
MEAS<X>
The automatic measurement number
from 1 to 8.
Return parameter
<NR3>
Returns the measurement for the
selected measurement number.
Example
:MEASUrement:MEAS3:SOUrce1 CH1
:MEASUrement:MEAS3:TYPe PK2PK
:MEASUrement:MEAS3:STATE ON
:MEASUrement:STATIstics:MODe ON
:MEASUrement:MEAS3:MINImum?
1.600E-02
Returns the minimum measurement result for
measurement number 3.
:MEASUrement:MEAS<X>:STDdev
Description
Returns the standard deviation for the selected
measurement from the last time the statistics were
reset. This is a statistics related command.
Syntax
:MEASUrement:MEAS<X>:STDdev?
Related
Commands
:MEASUrement:STATIstics:MODe
Parameter
MEAS<X>
The automatic measurement number
from 1 to 8.
Return parameter
<NR3>
Returns the measurement for the
selected measurement number.