COMMAND DETAILS
121
Syntax
:TRIGger:LEVel {TTL | ECL | SETTO50 | <NRf> | ?}
Related
commands
:TRIGger:TYPe
Parameter
<NRf>
Trigger level value.
TTL
Sets the trigger level to TTL.
ECL
Sets the trigger level to ECL.
SETTO50
Sets the trigger level to the User level
(50% by default).
Return parameter
<NR3>
Returns the trigger level.
Example1
:TRIGger:LEVel TTL
Sets the trigger to TTL.
Example2
:TRIGger:LEVel 3.30E-1
Sets the trigger level to 330mV/mA.
:TRIGger:HLEVel
Description
Sets or queries the high trigger level.
Note
Applicable for Rise and Fall/Pulse Runt triggers.
Syntax
:TRIGger:HLEVel {<NRf> | ?}
Related
commands
:TRIGger:TYPe
Parameter
<NRf>
High level value.
Return parameter
<NR3>
Returns the trigger high level.
Example
:TRIGger:HLEVel 3.30E-1
Sets the trigger high level to 330mV/mA.
:TRIGger:LLEVel
Description
Sets or queries the low trigger level.
Note
Applicable for Rise and Fall/Pulse Runt triggers.
Set
Query
Set
Query