GDS-1000B Programming Manual
98
Example
:MEASure:SOURce1 CH1
:MEASure:SOURce2 CH2
:MEASure:LRRDelay?
-8.89E-6
Select channel 1 and 2 as delay source1/2, and
then measure LRR.
:MEASure:LRFDelay
Description
Returns the delay between the first rising edge of
source1 and the last rising edge of source2.
Syntax
:MEASure:LRFDelay{?}
Related
Commands
:MEASure:SOURce<X>
Return parameter
<NR3>
Returns the delay.
Chan Off
Indicates the source channel is not
activated.
Note
Select the two source channels before entering this
command.
Example
:MEASure:SOURce1 CH1
:MEASure:SOURce2 CH2
:MEASure:LRFDelay?
-4.99E-6
Select channel 1 and 2 as delay source1/2, and
then measure LRF.
:MEASure:LFRDelay
Description
Returns the delay between the first falling edge of
source1 and the last rising edge of source2.
Syntax
:MEASure:LFRDelay{?}
Related
Commands
:MEASure:SOURce<X>
Query
Query

Table of Contents