EasyManua.ls Logo

GW Instek GDS-1102B - Acquire:filter:tracking; Acquire:recordlength; Acquire<X>:State

GW Instek GDS-1102B
168 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GDS-1000B Programming Manual
40
:ACQuire:FILTer:TRACking
Description
Turns filter tracking on/off or queries its state.
Note
This command is only applicable when the Digital
Filter app is installed. See the Digital Filter app
user manual for details.
Syntax
:ACQuire:FILTer:TRACking {ON|OFF|?}
Parameter/
Return parameter
OFF
Tracking off
ON
Tracking on
Example
:ACQuire:FILTer:TRACking ON
Turns filter tracking on.
:ACQuire<X>:STATe?
Description
Returns the status of waveform data.
Syntax
:ACQuire<X>:STATe?
Parameter
<X>
Channel number (1 to 4)
Return parameter
0
Raw data is not ready
1
Raw data is ready
Example
:ACQuire1:STATe?
0
Returns 0. Channel 1’s raw data is not ready.
Note: If the oscilloscope changes the acquisition status
from STOP to RUN, the status will be reset as zero.
:ACQuire:RECOrdlength
Description
Sets or queries the record length.
Syntax
:ACQuire:RECOrdlength {<NRf>| ?}
Parameter/Return
parameter
<NRf>
Record length. Settable record length:
(1e+3 | 1e+4 | 1e+5 | 1e+6 | 1e+7)
Set
Query
Query
Set
Query

Table of Contents

Related product manuals