EasyManua.ls Logo

GW Instek GDS-1102B - Measure:max; Measure:min

GW Instek GDS-1102B
168 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
COMMAND DETAILS
91
Example
:MEASure:SOURce1 CH1
:MEASure:LOW?
1.00E-0
Selects Channel 1, and then measures the low
current/voltage.
:MEASure:MAX
Description
Returns the maximum amplitude.
Syntax
:MEASure:MAX{?}
Related
Commands
:MEASure:SOURce<X>
Return parameter
<NR3>
Returns the maximum amplitude.
Chan Off
Indicates the source channel is not
activated.
Note
Before using this command, select the
measurement channel. See the example below.
Example
:MEASure:SOURce1 CH1
:MEASure:MAX?
1.90E-3
Selects Channel 1, and then measures the
maximum amplitude.
:MEASure:MIN
Description
Returns the minimum amplitude.
Syntax
:MEASure:MIN{?}
Related
Commands
:MEASure:SOURce<X>
Return parameter
<NR3>
Returns the minimum amplitude.
Chan Off
Indicates the source channel is not
activated.
Query
Query

Table of Contents

Related product manuals