EasyManua.ls Logo

GW Instek GDS-1102B - Measure:high; Measure:low

GW Instek GDS-1102B
168 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GDS-1000B Programming Manual
90
:MEASure:HIGH
Description
Returns the global high voltage/current.
Syntax
:MEASure:HIGH{?}
Related
Commands
:MEASure:SOURce<X>
Return parameter
<NR3>
Returns the high value.
Chan Off
Indicates the source channel is not
activated.
Note
Before using this command, select the
measurement channel. See the example below.
Example
:MEASure:SOURce1 CH1
:MEASure:HIGH?
3.68E-3
Selects Channel 1, and then measures the high
voltage/current.
:MEASure:LOW
Description
Returns the global low voltage/current.
Syntax
:MEASure:LOW{?}
Related
Commands
:MEASure:SOURce<X>
Return parameter
<NR3>
Returns the global low value.
Chan Off
Indicates the source channel is not
activated.
Note
Before using this command, select the
measurement channel. See the example below.
Query
Query

Table of Contents