REMOTE CONTROL
187
4 Execution
Error
16
An execution error.
5 Command
Error
32
A command syntax error.
6 Not used
64
Not used, return 0.
7 Power On
128
This bit is set if the power supply
has been reset from the last time
you read the event register.
The following will clear the standard event register:
The *CLS command is executed.
•The *ESR? command is used to query the event register.
The following will clear the standard event enable register.
The *ESE command is executed.
Status Byte Register Commands
Command
*SRE <Allowed values>
Function
Service request enable register (SRER) command
that writes a binary weighed value which
determines which events in the status byte
register are enabled.
Function
Allowed
values
Decimal vales: 0255
Example
*SRE 7
Returns the SRER setting (0000 0111)
Command
*SRE?
Function
Queries the status byte enable register. This
command returns a binary-weighted decimal
number that indicates which bits are set in the
status byte register. The range is from 0~255.