REMOTE CONTROL
191
Errors
Error Message
Errors are stored in a first in-first out (FIFO) order. The first error
message that is returned is the first error message that was stored.
When an error is read it is also cleared from the queue.
•If there are more than 10 errors produced the last error in the queue is
replaced with “Que overflow”. Unless the error queue is cleared, no
more errors can be written to the error queue. If there are no errors in
the error queue, the instrument will return “No error”.
•To clear the error queue, you can use the :SYSTem:CLEar command
or cycle the power. When you read a message from the error queue
that message will be cleared from the error queue. Using the *RST
command to reset the instrument does not clear the error queue.
•Remote control instructions can be used to clear the error queue. See
the instructions listed in the previous chapter for details.
Command Errors
-440 Query unterminated after indenite
-430 Response
-420 Query deadlocked
-410 Query unterminated
-363 Query interrupted
-350 Input buffer overrun
-330 Queue overow
-314 Self-test failed
-315 Save/recall memory lost
-260 Configuration memory lost
-241 Expression error
-230 Hardware missing