EasyManua.ls Logo

Handicare PUMA 40 - Service Manual

Handicare PUMA 40
138 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
SERVICE MANUAL
EN

Table of Contents

Other manuals for Handicare PUMA 40

Questions and Answers

  • M
    Monica HarrisSep 12, 2025
    What does Right Indicator Lamp Failed mean on my Handicare Scooter?
    • C
      conleytamaraSep 12, 2025
      A 'Right Indicator Lamp Failed' on your Handicare Scooter means that the control system has detected a failure in either of the Indicator Circuits. Check the indicator bulbs, cables, and connections to the control system. If the error persists after the above checks have been made, then the ISM may be defective.
  • L
    Laura MathewsSep 8, 2025
    What does it mean if the ON/OFF light flashes twice on my Handicare Scooter?
    • J
      jenniferwaltonSep 8, 2025
      If the ON/OFF light flashes twice on your Handicare Scooter, it indicates a DX accessory fault. The module displays a fault as a result of a programming fault or short circuit and/or overload of the accessories.
  • A
    Amanda WebbSep 6, 2025
    What does it mean when the ON/OFF light flashes 6 times on my Handicare Scooter?
    • D
      Daniel SmithSep 6, 2025
      If the ON/OFF light flashes 6 times on your Handicare Scooter, it indicates that the m2 parking brake connection is faulty or loose, or the m2 parking brake is defective. Correct the connection if it's faulty or loose, or replace the drive motor if the parking brake is defective.
  • T
    Thomas EnglishAug 28, 2025
    What causes an M1 Brake Error on my Handicare PUMA 40 Wheelchair?
    • S
      Sonya WilliamsAug 28, 2025
      An M1 Brake Error indicates a problem in the solenoid brakes or their connections. Check the solenoid brakes, cables, and connections to the control system. If the error persists, the Power Module may be defective.
  • L
    Luis SmithAug 28, 2025
    What causes an M2 Brake Error on my Handicare PUMA 40 Wheelchair?
    • J
      Jesse FowlerAug 28, 2025
      An M2 Brake Error indicates a problem in the solenoid brakes or their connections. Check the solenoid brakes, cables, and connections to the control system. If the error persists, the Power Module may be defective.
  • H
    huntkellyAug 28, 2025
    What does the M1 Motor Error mean on my Handicare PUMA 40 Wheelchair?
    • A
      Alexandra AllenAug 28, 2025
      The M1 Motor Error indicates that the control system has detected that a motor has become disconnected. Check the motors, cables, and connections to the control system. If the error persists after checking, the Power Module may be defective.
  • A
    Angelica DavisSep 23, 2025
    What does PM Memory Error mean on Handicare Mobility Aid?
    • M
      matthewmartinezSep 23, 2025
      If your Handicare Mobility Aid displays 'PM Memory Error', it is a specific Power Module-based trip. Check all cables and connections. Using the R-net PC Programmer, re-program the control system with either the most current specific program file for the wheelchair or the manufacturer's original programming file. If the trip persists, the PM may be defective.
  • J
    Joel ManningSep 15, 2025
    What does it mean if ON/OFF light flashes 4 times on Handicare Mobility Aid?
    • J
      Joe WigginsSep 15, 2025
      If the ON/OFF light on your Handicare Mobility Aid flashes 4 times, the cause may be related to the m2 motor. The m2 motor connection could be faulty or loose, so check the connection and secure it. The m2 motor itself might be defective and need replacement. There might also be a short circuit in the m2 motor connections, requiring correction. Additionally, a defective power module output could be the cause. Check the motor for any defects, as a power module output typically becomes defective due to a motor issue. Replace the power module only if the fault occurs immediately upon switching the unit on. If the fault occurs when you start driving, the entire left-hand motor circuit is likely causing the fault, in which case you should replace the entire drive unit concerned.
  • J
    John HarrisSep 18, 2025
    What causes M1 Brake Error on Handicare Mobility Aid?
    • O
      Oscar RobbinsSep 18, 2025
      If your Handicare Mobility Aid displays an 'M1 Brake Error', it indicates the control system has detected a problem in the solenoid brakes or their connections. Check the solenoid brakes, cables, and connections to the control system. If the error persists after these checks, the Power Module may be defective.
  • C
    Carrie JohnsonSep 8, 2025
    What does it mean if the ON/OFF light flashes once on my Handicare Scooter?
    • R
      Robert PierceSep 8, 2025
      If the ON/OFF light flashes once on your Handicare Scooter, it indicates a DX module fault. Replace one or more modules. The 'on/off' LED indicates the condition of the entire system.

Summary

Introduction

1.1 This manual

Explains the purpose and scope of the service manual for mechanics.

1.2 Identification of the product

Details the information found on the product's identification plate.

1.3 Symbols used in this manual

Explains warning and reference symbols used in the manual for clear communication.

Safety

2.1 Maximum User weight reduction

Lists Puma 20 options that reduce the maximum user weight on a Puma 40 carrier.

2.2 Personnel qualification

Specifies requirements for service technicians performing repairs.

2.3 Cautions and warning statements

Outlines general safety information, temperature warnings, and interference precautions.

2.4 Used decals on the wheelchair

Illustrates and explains the meaning of various decals found on the wheelchair.

Tools, Parts, and Components

3.1 Tools

Lists mechanical tools required for settings and maintenance tasks.

3.2 Tools electronics

Lists tools required for electronic settings and programming.

Spare Parts

4.1 Use of the parts lists

Explains how to use the parts lists for ordering replacement components.

4.2 Carrier and seat adjustments

Provides an overview of carrier and seat components with their corresponding page numbers.

4.3 Wiring and modules Shark

Lists wiring harnesses and electronic modules for the Shark system.

4.4 Wiring and modules R-net

Lists wiring harnesses and electronic modules for the R-net system.

4.5 Wiring and modules DX2

Lists wiring harnesses and electronic modules for the DX2 system.

Wiring and modules VR-2

Lists wiring harnesses and electronic modules for the VR-2 system.

Service Instructions

5.1 Maintenance plan

Details the regular maintenance schedule for the wheelchair, specifying tasks and responsibilities.

5.2 Assembly, replacement and adjustment instructions

Introduces instructions for assembly, replacement, and adjustment of wheelchair components.

Replacing Puma 20 motors

Guides on replacing Puma 20 motors, including compatibility and programming.

Converting Puma 20;40 FWD to RWD

Details the procedure for converting drive direction from FWD to RWD and vice versa.

Replacing carbon brushes

Provides steps for replacing carbon brushes in the motor assembly.

Replacing drive wheels

Explains the process for replacing the indoor/outdoor drive wheels.

Replacing castor wheels

Details the procedure for replacing the castor wheels.

Replacing the castor fork

Outlines steps for replacing the castor fork assembly.

Mounting direct access

Provides instructions for installing the direct access accessory.

Mounting attendant steering

Guides on installing the attendant steering system.

Replacing batteries

Details the procedure for replacing the wheelchair batteries, including safety precautions.

Replacing power module

Explains how to replace the power module unit.

Adjusting seating height

Details the process for adjusting the wheelchair's seating height.

Adjusting centre of gravity

Explains how to adjust the wheelchair's center of gravity for stability.

Adjusting electrical seat tilt

Explains how to adjust the electrical seat tilt angles.

Seat tilt configuration tables

Provides tables detailing seat tilt limitations based on various configurations.

Guidelines to solve shimmy problems

Offers troubleshooting steps for castor wheel shimmy issues.

Troubleshooting

6.1 Shark faultfinding table

Lists common faults and remedies for the Shark system.

6.2 R-net faultfinding table

Lists common faults and remedies for the R-net system.

6.3 DX2 faultfinding table

Lists common faults and remedies for the DX2 system.

DX2-ACT Flash Code List

Lists flash codes and their meanings for the DX2 Actuator Module.

6.4 VR-2 faultfinding table

Lists common faults and remedies for the VR-2 system.

Technical Product Information

7.1 CE Declaration and standards

Lists relevant CE directives and standards the product complies with.

7.2 Technical information

Provides general technical specifications like battery details and dimensions.

Model Specifications

Lists technical specifications for various Puma models (40/20, Sedeo Pro+/Pro/Lite/EasyFit).

7.3 Electrical diagrams

Shows system diagrams for various control modules like Shark, DX2, R-net, VR-2.

Shark Electrical Diagrams

Shows system diagrams for Shark control modules.

DX2 Electrical Diagrams

Shows system diagrams for DX2 control modules.

R-net Electrical Diagrams

Shows system diagrams for R-net control modules.

VR-2 Electrical Diagrams

Shows system diagrams for VR-2 control modules.

Warranty

8.1 Definitions of terms

Provides definitions for terms used in the warranty section.

8.2 Warranty period table

Outlines the warranty periods for different parts and components of the power wheelchairs.

Handicare PUMA 40 Specifications

General IconGeneral
BrandHandicare
ModelPUMA 40
CategoryWheelchair
LanguageEnglish

Related product manuals