60
NSLx Requests the number of logged samples.
The command parameter (1 char):
E ‑ the request is for EC range
R ‑ the request is for Resistivity range
T ‑ the request is for TDS range
N ‑ the requests is for NaCl range
U ‑ the requests is for USP range
The answer string is the number of samples (4 chars)
ULS Requests information about the status of the USP log reports.
The answer string contains:
• The number of reports (3 chars)
For every USP report the following information are sent:
• Report ID (3 chars)
• Log time (12 chars)
LLS Requests information about the number of lots
The answer string contains:
• The number of lots (3 chars)
For every lot the following information are sent:
• Lot ID (3 chars)
• Log time (12 chars)
• Lot type (1 char): 0 ‑ EC; 1 ‑ Resistivity; 2 ‑ TDS; 3 ‑ NaCl
GLDxxx Requests all the records for the lot with ID = xxx
The answer string contains:
Lot header data:
• Logging interval (1 s) (5 chars)
• Temperature source (1 char)
• Temperature compensation mode (1 char)
• Reference temperature (2 chars)
PAR Requests the setup parameters setting.
The answer string contains:
Number of profiles (2 chars)
Current profile ID (2 chars)
Number of languages (2 chars)
PC INTERFACE

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments HI98192 and is the answer not in the manual?

Hanna Instruments HI98192 Specifications

General IconGeneral
BrandHanna Instruments
ModelHI98192
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals