Hanna Instruments HI98192 - Page 61

Hanna Instruments HI98192
68 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
61
• Temperature coefficient (6 chars)
• Offset factor (6 chars)
• TDS factor (6 chars) (only for TDS lots)
• Salinity coefficient (7 chars) (only for salinity lots)
• Profile ID (4 chars)
• Profile creation time (12 chars)
• Lot start time (12 chars)
Lot record data:
• EC value (8 chars)
• EC unit (1 char): 0 ‑ µS; 1 ‑ mS
• EC over range flag (1 char): R, U, O
• Temperature reading, in °C (8 chars)
Next data is sent if the lot is not for EC range.
• Resistivity or TDS or Salinity (8 chars)
• Reading unit:
• Over range flag (1 char)
• Resistivity: 0 ‑ ohm; 1 ‑ Kohm; 2 ‑ Mohm
• TDS: 0 ‑ ppm; 1 ‑ g/L
• Salinity: 0 ‑ %; 1 ‑ ppt; 2 ‑ PSU
LODxxx Requests the log information for record number “xxx” on range “r
• “r” is E for EC, R for resistivity, T for TDS and N for salinity
The answering string contains:
• The logged range (2 chars): 10 ‑ EC, 11 ‑ Resistivity, 12 ‑ TDS, 13 ‑ USP,
14 ‑ NaCl %, 15 ‑ Salinity Seawater, 16 ‑ Salinity PSU
For EC, Resistivity, TDS and salinity ranges:
• EC reading (8 chars)
• EC unit (1 char)
• Temperature source (1 char)
• Temperature compensation mode (1 char)
• Reference temperature (2 chars)
• Temperature coefficient (6 chars)
• Cell Constant (7 chars)
• The nearest calibration standard (7 chars)
PC INTERFACE