EasyManuals Logo

Hanna Instruments HI98192 User Manual

Hanna Instruments HI98192
68 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #61 background imageLoading...
Page #61 background image
61
• Temperature coefficient (6 chars)
• Offset factor (6 chars)
• TDS factor (6 chars) (only for TDS lots)
• Salinity coefficient (7 chars) (only for salinity lots)
• Profile ID (4 chars)
• Profile creation time (12 chars)
• Lot start time (12 chars)
Lot record data:
• EC value (8 chars)
• EC unit (1 char): 0 ‑ µS; 1 ‑ mS
• EC over range flag (1 char): R, U, O
• Temperature reading, in °C (8 chars)
Next data is sent if the lot is not for EC range.
• Resistivity or TDS or Salinity (8 chars)
• Reading unit:
• Over range flag (1 char)
• Resistivity: 0 ‑ ohm; 1 ‑ Kohm; 2 ‑ Mohm
• TDS: 0 ‑ ppm; 1 ‑ g/L
• Salinity: 0 ‑ %; 1 ‑ ppt; 2 ‑ PSU
LODxxx Requests the log information for record number “xxx” on range “r
• “r” is E for EC, R for resistivity, T for TDS and N for salinity
The answering string contains:
• The logged range (2 chars): 10 ‑ EC, 11 ‑ Resistivity, 12 ‑ TDS, 13 ‑ USP,
14 ‑ NaCl %, 15 ‑ Salinity Seawater, 16 ‑ Salinity PSU
For EC, Resistivity, TDS and salinity ranges:
• EC reading (8 chars)
• EC unit (1 char)
• Temperature source (1 char)
• Temperature compensation mode (1 char)
• Reference temperature (2 chars)
• Temperature coefficient (6 chars)
• Cell Constant (7 chars)
• The nearest calibration standard (7 chars)
PC INTERFACE

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments HI98192 and is the answer not in the manual?

Hanna Instruments HI98192 Specifications

General IconGeneral
BrandHanna Instruments
ModelHI98192
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals