Transducer Initialization
Gladiator Acoustic Switch Series
24
Error 01 - Error 11
• If the LCD is displaying ‘Error 01 or Error 11 after
installing a single transducer or full system you may
need to re-initialize a transducer.
• If your system is displaying Error 01 or Error 11 after
the system has been working correctly it is possible
there is a hardware problem.
• If Comm Err is displayed while navigating menus
check your wiring, terminals, junction boxes and
transducers for damage or connection problems.
Initializing a Transducer
• If you need to re-initialize the transducer address
you can run this program.
• You will have the option to clear each & initialize
transducer address.
• First - Disconnect Transducer 2
• Press and hold CAL to force the unit to open the
‘unlock’ menu.
• Navigate to the Initialize TX menu as per below:
After selecting ‘Yes’, you will see one of these
messages:
T’ducer1 Initialize?
• Transducer 1 is detected and has not been initialized.
• Press CAL to edit, press UP until you see ‘Yes’.
• Press CAL to select.
• Re-connect Transducer 2 and press RUN several
times to re-activate the unit.
T’ducer 1 Ready
• Press CAL to edit, press UP twice to display ‘Clear?
Yes’.
• Press CAL.
• The unit will cycle to T’ducer 2 displaying ‘Ready’.
• Re-connect Transducer 2 and press RUN several
times to re-activate the unit.
Plz connect Transducer 1
• Press RUN firmly until it moves to 2nd Transducer.
• Display should read ‘T’ducer 2 Ready’ (if this does
not occur, skip to below note).
• Press CAL to edit, press UP twice to display ‘Clear?
Yes’. Press CAL.
• The unit will not have returned to the advanced menu.
• Press UP and re-enter TX initialize.
• You should now be prompted with T’ducer1 Initialize,
refer to relevant steps.
*Note*:
• If the unit menu does not scroll to ‘T’ducer 2 Ready’
you will need to perform a ‘Load Defaults’ routine
(located in Advanced Menu). This will prompt you
to load defaults and confirm selection followed by
resetting the sensor - do not reset the sensors.
• After the load default press RUN to re-start the unit.
Next, perform a power cycle of the entire system.
• When prompted to connect Wire Tx 1, Press CAL.
When prompted to connect Transducer 1, press RUN.
• ‘T’ducer 2 Ready’ should now be displayed.
Press UP arrow to display ‘Clear’. Press CAL.
The unit will now re-start.
• Follow on-screen instructions to complete setup.
Initilize TX
Unlock 0
CAL
x3
Advanced
CAL
CAL
*edit* Yes
CAL