1 Unpacking and Setting up
1–14 Edition October 1998
■
Removing transport safeguards for the
scanning head
Remove the 4 transport safeguards for the
scanning head by firstly unscrewing the 4
screws using a 10 mm wrench and then
removing the shims still attached with
adhesive tape.
Make sure that you do not let the screws
and shims fall into the unit.
Now pull the steel cable forwards and
remove the plastic block.
■
Removing the transport safeguard for the
focus motor (1) (for TANGO only, not for
TANGO XL) and the edge protection (2)
Remove the transport safeguard by pulling
it upwards. It ensures that the focus motor
does not become jammed in transport posi-
tion, thus preventing errors occurring dur-
ing installation.
Subsequently, remove the transparent edge
protection. For this purpose, you have to lift
off the lamp cover.
1
2