Appendix
Page 62
heyer_LUVAR_UserManual_EN-3_CRev-1
Adjustable start pressure min. 4 hPa
Adjustable ramp time Off; 5, 10, 15 ... 60 minutes
Technical data for power supply unit
Model name DA-80A24
Weight 0.34 kg
Electrical protection class Protection class II
Input voltage 100 - 240 V AC
Input current max. 2.0 A
eff
Input frequency 50 - 60 Hz
Output voltage 24 V DC ±1.2 V
Output current max. 3.33 A
Technical data for respiratory humidifier H100
Dimensions W × D × H (integrated in
device)
11.5 x 15 x 10 cm (10.5 x 15 x 10 cm)
Weight 0.42 kg
Air outlet port diameter 22 mm, conical (ISO 5356-1)
Input voltage 24 V DC
Max. power consumption 30 W
Max. temperature of gas delivered 43 °C
Heating level, adjustable Off, 1 to 5
Water tank capacity ≤ 300 ml
Operating time between refills at heating
level 5
8 hours
Humidifier output as per ISO 80601-2-74 >12.5 mg/l
Stability of static pressure (long-term accuracy) at 10 hPa as per
ISO 80601-2-70
Maximum deviation from set pressure for LUVAR A/STA without and with humidifi-
er, with unheated breathing tube: Δp ≤ 0.39 hPa.