Appendix
Page 71
heyer_LUVAR_UserManual_EN-3_CRev-1
Interference
emission
Compliance Electromagnetic environment - guideline
RF emissions as
per CISPR 11
Group 1 The device uses RF energy only for
its internal function. Therefore, its RF
emissions are very low and unlikely to
cause interference to nearby electronic
equipment.
RF emissions as
per CISPR 11
Class B The device can be used in all facilities,
including domestic facilities and facil-
ities directly connected to the public
power supply network.
Emissions of har-
monics according
to IEC 61000-3-2
Class A
Emissions of volt-
age fluctuations/
flicker according
to IEC 61000-3-3
Compliant
Electromagnetic immunity
Guidelines and manufacturer’s declaration – electromagnetic immunity
The LUVAR respiratory therapy device with or without the H100 respiratory hu-
midifier is intended for use in domestic settings as well as in appropriate clinical
settings. Device users must ensure that it is used in such a setting. When used in a
residential setting (for which Class B is typically required by CISPR 11), this device
may not provide adequate protection from radio communication services. The user
may need to take remedial action such as relocating or repositioning the device.