A n e s t h e s i a S y s t e m
Operator’s Manual
Rev. 0.2 Draft – 12/09
I
NNOVATION IN
D
ESIGN AND
T
ECHNOLOGY

Table of Contents

Questions and Answers

  • M
    Manuel HendersonAug 13, 2025
    How to fix a breathing system leak in Heyer Pasithec?
    • P
      Paul BrownAug 14, 2025
      A breathing system leak in Heyer Medical Equipment can occur for several reasons. Check if the APL valve is on and turn it off if it is. Ensure the canister is sealed well; reinstall or remove any natrium calcareousness grains at the joint. Inspect the corrugated tubing for breaks or loose connectors, and replace or reinstall as necessary. Finally, make sure all valves are tightened properly.
  • D
    Dr. Richard CraneAug 17, 2025
    What to do if the maximum pressure alarm sounds continuously on Heyer Medical Equipment?
    • T
      todd06Aug 17, 2025
      If the maximum pressure alarm sounds continuously on your Heyer Medical Equipment, several factors could be the cause. First, check the breathing system for any occlusions and adjust accordingly. Next, examine the patient for any respiratory tract occlusions. Also, verify that the maximum pressure setting isn't too low and readjust if needed. Finally, if ventilator parameters have changed, recalculate them.
  • T
    timothygreeneAug 21, 2025
    Why the AC indicator has no power, and ventilator does not work in Heyer Pasithec?
    • I
      Isabel JohnsonAug 21, 2025
      If the AC indicator has no power and the ventilator isn't working on your Heyer Medical Equipment, check the following: Ensure the power supply cable is properly plugged in. Verify that the power switch is turned on. Finally, inspect the fuse and replace it if it's burned out.
  • K
    Kevin CainAug 25, 2025
    What to do if minimum pressure alarm sounds continuously in Heyer Pasithec Medical Equipment?
    • J
      johnsoncharlesAug 25, 2025
      If the minimum pressure alarm sounds continuously on Heyer Medical Equipment, check the following: Inspect the breathing system for leaks. Reset the alarm settings if they are too high. Check the patient. Inspect the pressure sampling pipe.
  • A
    Andrew HenryAug 28, 2025
    Why my Heyer Medical Equipment ventilator stops suddenly and the alarm sounds?
    • A
      aromeroAug 28, 2025
      If your Heyer Medical Equipment ventilator stops suddenly, the indicator light turns off, and an alarm sounds, it indicates that the power supply has been interrupted. Use manual ventilation.
  • S
    Stephen BennettSep 1, 2025
    How to troubleshoot no indication from the airway pressure gauge on Heyer Pasithec Medical Equipment?
    • A
      Amanda VaughnSep 1, 2025
      If there is no indication from the airway pressure gauge on your Heyer Medical Equipment, check if the pressure sampling pipe is disconnected and reconnect it. Also, verify that the gas supply is not exhausted and replace it if necessary.
  • Y
    yatesanneSep 4, 2025
    What causes tidal volume readings to not display normally on Heyer Medical Equipment?
    • T
      troy82Sep 5, 2025
      If tidal volume readings are not displaying normally on your Heyer Medical Equipment, it could be due to a disconnected flow sensor. Reconnect the flow sensor. Alternatively, the sealing ring of the bellows base might be broken. Check the bellows and replace any broken parts.
  • R
    riveraderekSep 8, 2025
    Why is there excessive pressure caused by manual ventilation with Heyer Medical Equipment?
    • O
      oliviafritzSep 8, 2025
      Excessive pressure during manual ventilation with Heyer Medical Equipment can be caused by incorrect adjustment of the APL valve. Adjust it properly.

Summary

System Components Overview

Anesthetic System Components

Details the required monitoring devices, alarm systems, and protection devices for the anesthetic system.

Breathing System Description

Illustrates and labels the components of the breathing system used with the Pasithec.

Anesthetic Ventilator Details

Describes the anesthetic ventilator, its compliance, and monitoring conditions.

Anesthetic Ventilator Screen Layout

Explains the screen layout, including information, monitoring, and parameter setup areas.

Operating the HEYER Pasithec

System Startup Procedures

Step-by-step guide for connecting power, gas, and initiating the system self-test.

Leak and Safety Valve Testing

Procedures for performing manual, automatic leak tests, and safety valve tests.

Mechanical Ventilation Operations

Guide on starting and stopping mechanical ventilation, including auto-ventilation.

Manual Ventilation and Shutdown

Instructions for starting, stopping manual ventilation, and shutting down the system.

Pre-Use Check List and Tests

Pre-use Check List Procedure

Defines the schedule and steps for pre-operative checkout and daily checks.

Alarm System Testing

Detailed steps for testing various alarms, including O2 monitoring, volume, and pressure alarms.

Ventilator and Breathing System Tests

Procedures for testing the ventilator and breathing system components, including O2+ button.

Installation and Connection Procedures

Cleaning and Disinfecting Procedures

User Maintenance and Care

O2 Sensor Calibration

Detailed steps for calibrating the 21% O2 sensor and the importance of calibration.

Alarms and Troubleshooting

Technical Alarm Messages

Lists technical alarms, their causes, machine modes, operator actions, and required repairs.

Functional Alarm Conditions

Details functional alarms, their conditions, manual mode disable status, operator actions, and repairs.

Troubleshooting Common Issues

Provides a guide to diagnose and resolve common symptoms like breathing system leaks or pressure issues.

Specifications and Operation Theory

Heyer Pasithec Specifications

General IconGeneral
BrandHeyer
ModelPasithec
CategoryMedical Equipment
LanguageEnglish