Rev. 0.2 Draft – 12/09 HEYER Pasithec, Operator’s manual 69 / 106
8 User Maintenance
WARNING: To avoid fire:
Use the lubricant approved for anesthesia or O
2
equipments’ use.
Never oil or grease any anesthesia or O
2
equipment. In general, oils and
greases oxidize readily, and in the presence of O
2
are highly flammable.
All the covers or housings for the system use must be made of static-proof
material, as static material may cause fire.
WARNING: Follow sterilizing control and security stipulations because used equipment
may contain blood and body fluids.
WARNING: Movable components and detachable parts can cause injury. Use caution
when system components and parts are being moved or replaced.
WARNING: Shock and strong vibration during transportation can break the fragile glass
cover of the flowmeter.
WARNING: Disposal of waste or invalidated apparatus must be in accordance with local
laws.
8.1 Repair Policy
Do not use malfunctioning equipment. Make all necessary repairs, or have the equipment serviced by
an authorized HEYER Service Representative. After repair, test the equipment to ensure that it is
functioning properly and in accordance with the manufacturer’s published specifications.
To ensure full reliability, have all repairs and service done by an authorized representative. If this is
not possible, replacement and maintenance of parts in this manual should be performed by a
competent, trained individual with experience in Anesthesia Systems repair and appropriate testing
and calibration equipment.
CAUTION: No repair should ever be undertaken or attempted by anyone without proper
qualifications and equipment.
It is recommended that damaged parts be replaced with components manufactured or sold by
HEYER. After any repair work, test the unit to ensure it complies with the manufacturer’s published
specifications.
Contact the nearest HEYER Service Center for service assistance. In all cases, other than where our
warranty is applicable, repairs will be made at current list price for the replacement part(s) plus a
reasonable labor charge.
8.2 Maintaining Outline and Schedule
The following schedule is a recommended minimum standard based upon normal usage and
environmental conditions. Frequency of maintenance for the equipment should be higher if your actual
schedule is more than the minimum standard.