Rev. 0.2 Draft – 12/09 HEYER Pasithec, Operator’s manual 93 / 106
10.8.2 Ventilation Mode Settings
Ventilation mode Adjustable respiratory parameters
IPPV V
T
, Freq, I:E, T
P
, PEEP
PCV P
TARGET
, Freq, I:E, PEEP, T
SLOPE
PS Freq
MIN
, PEEP, P, Trigger, T
SLOPE
SIMV V
T
, Freq, Tinsp, T
P
, PEEP, P, Trigger, T
SLOPE
Manual mode ----
10.8.3 Ventilating Parameters Settings
Item Range Resolution Default Remark
V
T
20~1500 ml 10 ml 120 ml (Child);
600 ml (Adult)
Unavailable for SIMV
& IPPV mode.
Freq. 4~100 bpm 1 bpm 20 bpm (Child)
8 bpm (Adult)
Unavailable for SIMV,
PCV & IPPV mode.
I:E 4:1
3.5:1 3:1 2.5:1 2:1
1.5:1
1:1 1:1.5 1:2 1:2.5
1:3
1:3.5 1:4 1:4.5 1:5
1:5.5
1:6 1:6.5 1:7 1:7.5
1:8
0.5 1:2 Unavailable for SIMV
& IPPV mode.
T
INSP
0.2~5.0sec 0.1 sec 1 sec (Child)
2 sec (Adult)
Only available for
SIMV mode.
PEEP Off, 3~30 cmH
2
O 1 cmH
2
O Off Unavailable for SIMV,
PCV, PS & IPPV
mode.
Freq
MIN
2~60 bpm 1 bpm 4 bpm (Child);
2 bpm (Adult)
Only available for PS
mode.
T
P
Off, 5~60% 5% 10% Only available for
IPPV and SIMV
mode.
Trigger 1~15 L/min 1 L/min 2 L/min (Child);
3 L/min (Adult)
Only available for
SIMV, PS mode.
P
TARGET
5~70 cmH
2
O 1 cmH
2
O 10 cmH
2
O (Child);
20 cmH
2
O (Adult)
Only available for
PCV mode.
P 3~50 cmH
2
O 1 cmH
2
O 5 cmH
2
O Only available for
SIMV, and PS mode.
T
SLOPE
0~2 sec 0.1 sec 0.5 sec Available for PCV,
SIMV, and PS mode.
10.8.4 Gas Dynamics Performance
Gas source Anesthetic system
Gas component O
2
Rating supply pressure 400 kPa
Inlet pressure range 280 kPa ~ 600 kPa
Flow valve range 5~75 L/min
Output Pressure range: 0 ~ 6 kPa; Flow range: 0 ~ 75 L/min