8.3.4 Calibrate Agents
Step 1
Turn the knob to select “Agents Calibration” in the “Calibrate” submenu.
Step 2
Press the knob and a new window appears on the right side.
Perform the operation according to the prompt on the screen.
Disconnect the Gas Sensor from the Y-piece and place it in the air for no less than 15 minutes.
Remove the airway adapter (or a new one) until the light turns green.
Wait 1 minute till the State-Area turns green (OK) and select NEXT.