Rev. 0.2 Draft – 12/09 HEYER Pasithec, Operator’s manual 75 / 106
Step 3
Press NEXT to perform the calibration.
In the process of calibration, the word “Calibrating” displays on the screen.
“Calibration successful” will be displayed on the window after the calibration is completed successfully.
Reinstall the Gas Sensor to the breathing system. Select “Finish” to complete the calibration.