7.2 Cleanable Breathing System Components
Corrugated tubing (contacted with
patient), face mask, Y-piece
connector, L-piece, reservoir bag
Designed for using only once, no need to sterilize.
The waste should be recovered.
When replacing, ensure new products have the same
specifications.
Corrugated tubing and bag
(repetitious)
Wash to sterilize
7.3 Absorber Circle
Component
Wiping
method
Soap water
Cleaning
method
Soap water
Chemicals
Cidex
Sterilization
Steam
autoclave
Maximum
temperature
F/°C)
Absorber canister yes yes yes no
Airway pressure gauge yes no no no
APL valve yes yes yes no
Valve cover and valve
patch
yes yes yes no
Absorber circuit
integration (without the
above mentioned)
yes yes yes yes 273°F/134°C